ULTRA-COAT MICRO EJECT KERRISON 40D UP 8 3MM
Report
- Report Number
- 2523190-2016-00179
- Event Type
- Malfunction
- Date Received
- October 14, 2016
- Report Date
- September 16, 2016
- Manufacturer
- INTEGRA YORK, PA INC.
- Product Code
- HAE
- PMA / PMN Number
- K150428
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
10/17/16 INTEGRA INVESTIGATION COMPLETED. METHOD: FAILURE ANALYSIS, DEVICE HISTORY EVALUATION. RESULTS: FAILURE ANALYSIS - FAILURE ANALYSIS CANNOT BE PERFORMED BASED ON THE LACK OF INFORMATION PROVIDED BY THE CUSTOMER. THE RONGEURS WERE NOT RETURNED FOR FURTHER EVALUATION. DEVICE HISTORY EVALUATION - NONCONFORMING PRODUCT REPORT / NONCONFORMING MATERIAL REPORT HISTORY: THERE IS NO APPLICABLE NONCONFORMING PRODUCT REPORT / NONCONFORMING MATERIAL REPORT HISTORY. VARIANCE AUTHORIZATION / DEVIATION HISTORY: NONE. ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY: THERE IS NO APPLICABLE ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY. CORRECTIVE ACTION PREVENTIVE ACTION HISTORY/CORRECTION HISTORY: NONE. HEALTH HAZARD EVALUATION HISTORY: NONE. CONCLUSION: ROOT CAUSE CANNOT BE DETERMINED DUE TO THE LACK OF INFORMATION RECEIVED TO PERFORM A COMPLETE INVESTIGATION. THE RONGEURS WERE NOT RETURNED FOR FURTHER EVALUATION.
THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION
CUSTOMER INITIALLY REPORTS SCREWS FELL OUT OF 2 KERRISONS. DOCTORS ARE COMPLAINING THEY ARE NOT SHARP, THEY ARE NOT CUTTING, THEY ARE TEARING. ON (B)(6) 2016 NO HARM DONE. ON 9/30/16 CUSTOMER REPORTS SHE HAS NO SPECIFIC INFORMATION FOR AN EVENT EXCEPT ONE OF THE SCREWS DID FALL INTO A PATIENT WOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682454 | ULTRA-COAT MICRO EJECT KERRISON 40D UP 8 3MM | ULTRA-COAT | HAE | INTEGRA YORK, PA INC. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |