FDA Adverse Event Malfunction Summary report: N

TERUMO ADVANCED PERFUSION SYSTEM 1

MDR report key: 6026365 · Received October 13, 2016

Report

Report Number
1828100-2016-00674
Event Type
Malfunction
Date Received
October 13, 2016
Date of Event
September 19, 2016
Report Date
December 30, 2016
Manufacturer
TERUMO CARDIOVASCULAR SYSTEM CORP.
Product Code
DTQ
PMA / PMN Number
K151349
Removal / Correction Number
1828100-06/05/2007-005-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION IS IN PROGRESS, BUT NOT YET CONCLUDED. THE SYSTEM WAS LAST SERVICED INCLUDING SOFTWARE WAS UPGRADE PER NOTICE OF FIELD CORRECTION (NFC) ON 16-SEP-2016.

Additional Manufacturer Narrative · 1

DURING LABORATORY ANALYSIS, THE PRODUCT SURVEILLANCE TECHNICIAN (PST) OBSERVED ¿SYSTEM COMPUTER NEEDS SERVICE¿ MESSAGE DURING BOOT PROCESS, DUPLICATING THE COMPLAINT CONDITION. REPLACEMENT OF CUSTOMER DISK-ON-CHIP WITH LAB USE ONLY (LUO) DISK-ON-CHIP RESTORED CENTRAL CONTROL MONITOR (CCM)'S FUNCTIONALITY. THE PRODUCT WILL BE SENT TO SERVICE TO BE BROUGHT TO MANUFACTURERS SPECIFICATIONS BEFORE BEING RETURNED TO THE CUSTOMER. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 1

THE REPORTED COMPLAINT WAS CONFIRMED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE "SYSTEM COMPUTER NEEDS SERVICE" ERROR DISPLAYED AT POWER UP ON THE CENTRAL CONTROL MONITOR (CCM). THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
677958 TERUMO ADVANCED PERFUSION SYSTEM 1 CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 DTQ TERUMO CARDIOVASCULAR SYSTEM CORP. 802100

Patients

Seq Age Sex Outcome Treatment
1