FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 6026280 · Received October 13, 2016

Report

Report Number
2531779-2016-28390
Event Type
Malfunction
Date Received
October 13, 2016
Report Date
September 20, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/02/2016 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED MULTIPLE CALL SERVICE 006 ALARMS. THE PUMP POWERED ON TO A CALL SERVICE 006 ALARM. THE PUMP COVER WAS REMOVED TO FIND NO INTERMITTENT CONDITIONS ON THE PRINTED CIRCUIT BOARD. EVALUATION REVEALED THAT THE EEPROM COMPONENT HAD FAILED.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM ISSUE (CS 006-0001 ALARM). THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
677557 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 13 YR