FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 6025513 · Received October 13, 2016

Report

Report Number
2531779-2016-28353
Event Type
Malfunction
Date Received
October 13, 2016
Report Date
September 21, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1; DATE OF SUBMISSION: 11/2/2016. THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/03/2016 WITH THE FOLLOWING FINDINGS. CRACKED BATTERY COMPARTMENT FROM THREADS TO CASE SEAL LEFT OF GRIP PAD. MOISTURE OBSERVED IN BATTERY COMPARTMENT.. LEAK TEST FAILED DUE TO BATTERY COMPARTMENT LEAK OPENED PUMP, MOISTURE DAMAGE OBSERVED ON VIBRATION MOTOR, AND CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
676967 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1