ANIMAS VIBE
Report
- Report Number
- 2531779-2016-28353
- Event Type
- Malfunction
- Date Received
- October 13, 2016
- Report Date
- September 21, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1; DATE OF SUBMISSION: 11/2/2016. THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/03/2016 WITH THE FOLLOWING FINDINGS. CRACKED BATTERY COMPARTMENT FROM THREADS TO CASE SEAL LEFT OF GRIP PAD. MOISTURE OBSERVED IN BATTERY COMPARTMENT.. LEAK TEST FAILED DUE TO BATTERY COMPARTMENT LEAK OPENED PUMP, MOISTURE DAMAGE OBSERVED ON VIBRATION MOTOR, AND CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.