FDA Adverse Event Injury Summary report: N

ECHELON-10, 45 TIP SHAPED MICRO CATHETER

MDR report key: 602529 · Received May 13, 2005

Report

Report Number
2029214-2005-00032
Event Type
Injury
Date Received
May 13, 2005
Date of Event
April 13, 2005
Report Date
May 13, 2005
Manufacturer
MICRO THERAPEUTICS, INC.
Product Code
KRA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

COILING OF CEREBRAL ANEURYSM IN A 40 YRS OLD GRADE 1 SAH 2 DAYS POST ICTUS. ON INSPECTION OF THE DEVICE, TIP SHAPED OF THE CATHETER WAS NOTED TO BE SHAPED-25' RATHETER THAN 45'. ENTRY INTO ANEURYSM STRAIGHT FORWARD WITH MICROWIRE. WHILE ADVANCING THE MICRO CATHETER INTO THE ANEURYSM, THE PT SUDDENLY BECAME ACUTELY HYPERTENSIVE. CHECK ANGIOGRAPHY REVEALED RUPTURE OF THE ANEURYSM. A COIL WAS RAPIDLY DEPLOYED INTO THE MICRO CATHETER BUT WENT STRAIGHT THROUGH THE ANEURYSM (AS THE TRUE TIP OF THE CATHETER WHICH WAS NOT MARKED WAS RIGHT THOUGH THE ANEURYSM). ON REALIZING THAT THE DISTAL MARKER OF THE CATHETER WAS ABSENT, THE CATHETER WAS WITHDRAWN, AN ALTERNATE SELECT CATHETER WAS PLACED TO THE ANEURYSM AND THE ANEURYSM WAS RAPIDLY COILED. THE PT WAS NOT ANTICOAGULATED, CT SCAN SHOWED FURTHER FRESH BLOOD IN THE SYLVIAN FISSURE AND HYDROCEPHALUS. THE PT WAS EXTRUBATED AND HAS MADE A FULL RECOVERY WITH HYPERTENSIVE RX.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON-10, 45 TIP SHAPED MICRO CATHETER DELIVERY CATHETER KRA MICRO THERAPEUTICS, INC. 145-5091-150 1049055

Patients

Seq Age Sex Outcome Treatment
1 40 YR Life Threatening