PUMP IN STYLE ADVANCED - ON-THE-GO TOTE
Report
- Report Number
- 1419937-2016-00261
- Event Type
- Malfunction
- Date Received
- October 12, 2016
- Date of Event
- September 13, 2016
- Report Date
- September 13, 2016
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
A REPLACEMENT POWER ADAPTER WAS SENT TO THE CUSTOMER. A PRODUCT EVALUATION COMPLETED ON 10/10/2016 2016 FOUND THE TRANSFORMER HOUSING TO BE BREACHED. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING MONITORED FOR EFFECTIVENESS OF THE ABOVE MENTIONED CORRECTIVE ACTION.
ON (B)(6) 2016, THE CUSTOMER REPORTED TO A MEDELA CUSTOMER SERVICE REPRESENTATIVE THAT THE TRANSFORMER FOR HER PUMP IN STYLE ADVANCED BREAST PUMP BROKE IN HALF AND COMPLETELY SEPARATED. SHE SAID THAT SHE HEARD A LOUD POPPING AND SAW SMOKE COMING FROM THE ADAPTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 674095 | PUMP IN STYLE ADVANCED - ON-THE-GO TOTE | PUMP, BREAST, POWERED | HGX | MEDELA, INC. | 57063 / 9207010 | 5112 / REV M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |