FDA Adverse Event Injury Summary report: N

UNKNOWN

MDR report key: 6016392 · Received October 11, 2016

Report

Report Number
3002808486-2016-01225
Event Type
Injury
Date Received
October 11, 2016
Report Date
February 6, 2020
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 0

CONCLUSION CODE(S): APPROPRIATE TERM/CODE NOT AVAILABLE (4316) WAS SELECTED BECAUSE THE PREVIOUS INVESTIGATION REMAINS UNCHANGED BY THE ADDITIONAL INFORMATION. INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. PER QUALITY ENGINEERING REVIEW, THE ADDITIONAL INFORMATION PROVIDED FOR THIS COMPLAINT DOES NOT CHANGE THE PREVIOUS INVESTIGATION CONCLUSION. THEREFORE, NO NEW INVESTIGATION ACTIVITIES WILL BE CONDUCTED AT THIS TIME. COOK WILL REOPEN THE INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56 THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MFR SITE: NAME AND ADDRESS FOR IMPORTER SITE: (B)(4). ADDITIONAL INFORMATION: INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'CELECT, FRACTURE, MIGRATION (DISPLACED), ORGAN PERFORATION, ABDOMEN PAIN (DEATH)'. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. UNKNOWN IF THE REPORTED PATIENT DEATH IS RELATED TO THE FILTER. MANIPULATION IN THE AREA OF THE FILTER IMPLANT MAY CAUSE MIGRATION OR CONTRIBUTE TO CHANGES IN THE FILTER CONFIGURATION AND PLACEMENT. VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. FILTER FRACTURE IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. FRACTURE OF A FILTER LEG CAN BE DUE TO REPETITIVE MOTION ON A FILTER LEG IN AN UNUSUAL STRESSED POSITION. AMONG OTHER CAUSES, FILTER FRACTURE MAY BE ASSOCIATED WITH A FILTER LEG PERFORATING THE IVC, A FILTER LEG BEING CAUGHT IN A SIDE BRANCH (E.G. RENAL VEIN), EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G. A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND / OR PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. IT HAS BEEN REPORTED THAT RETRIEVAL OF A FRACTURED FILTER OR FILTER FRAGMENTS USING ENDOVASCULAR TECHNIQUES IS POSSIBLE. FRACTURE OF THE WIRE IS A KNOWN RISK IN RELATION TO AN IMPLANTED FILTER AND REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE. IT IS KNOWN FROM THE PUBLISHED SCIENTIFIC LITERATURE THAT A FILTER FRAGMENT EMBOLIZED INTO THE HEART OR LUNG MAY BE SAFELY RETRIEVED. UNKNOWN IF THE REPORTED ABDOMEN PAIN IS RELATED TO THE FILTER. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

PER DEATH CERTIFICATE: CAUSE OF DEATH: SEVERE SEPSIS, STAGE IV DECUBITUS ULCER, PNEUMONIA WITH RESPIRATORY FAILURE, RECENT SUBARACHNOID HEMORRHAGE (DUE TO ANEURYSM)

Additional Manufacturer Narrative · 1

(B)(4). SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K090140. (B)(4). INVESTIGATION IS STILL IN PROGRESS.

Additional Manufacturer Narrative · 1

MANUFACTURER REF# (B)(4). CATALOG#: UNKNOWN BUT REFERRED TO AS A COOK CELECT FILTER. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K090140. (B)(4). INFORMATION REGARDING THE EVENT HAS NOT BEEN PROVIDED. IT HAS NOT BEEN POSSIBLE TO INVESTIGATE THIS EVENT BASED ON THE LIMITED INFORMATION, AND WE ARE CLOSING THE REPORT UNTIL FURTHER INFORMATION IS RECEIVED. IF ADDITIONAL INFORMATION IS RECEIVED, THE REPORT WILL BE RE-OPENED FOR FURTHER INVESTIGATION.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016032. (B)(4). THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2009". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2009". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION RECEIVED ON 28FEB2018 AS FOLLOWS: PT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2009 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO DEEP VEIN THROMBOSIS AND CONTRAINDICATION TO ANTICOAGULATION. PT ALLEGES MIGRATION, FRACTURE, DISPLACED. PT IS REPORTED TO BE EXPIRED WITH NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
667749 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening