FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM PROFESSIONAL CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 6015374 · Received October 10, 2016

Report

Report Number
3004753838-2016-72469
Event Type
Malfunction
Date Received
October 10, 2016
Date of Event
September 16, 2016
Report Date
September 16, 2016
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000064
PMA / PMN Number
P120005
Removal / Correction Number
3004753838-02/29/16-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016 THAT ON (B)(6) 2016 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE REPORTED NO AUDIO OUTPUT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

THE RECEIVER WAS RETURNED FOR EVALUATION. THE RECEIVER WAS VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. A "TRY IT" MANUAL TEST WAS PERFORMED AND SPEAKER DID NOT SOUND. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
664964 DEXCOM G4 PLATINUM PROFESSIONAL CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM, INC. MT20649 5203485 00386270000064

Patients

Seq Age Sex Outcome Treatment
1