FDA Adverse Event Injury Summary report: N

LINX REFLUX MANAGEMENT SYSTEM

MDR report key: 6014013 · Received October 10, 2016

Report

Report Number
3008766073-2016-00070
Event Type
Injury
Date Received
October 10, 2016
Date of Event
August 31, 2016
Report Date
October 25, 2016
Manufacturer
TORAX MEDICAL, INC.
Product Code
LEI
UDI-DI
00855106005035
PMA / PMN Number
P100049
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

-DEVICE ANALYSIS CONDUCTED BY TORAX MEDICAL ENGINEERING AFTER PRODUCT RECEIPT. GROSS ANALYSIS REVEALED NO DEVICE ANOMALIES ATYPICAL FROM AN EXPLANTED DEVICE. MICROSCOPIC ANALYSIS REVEALED ALL DEVICE COMPONENTS AND ASSEMBLY EXHIBITED NORMAL CHARACTERISTICS OF AN EXPLANTED DEVICE. FORCE TESTING, LENGTH TESTING, AND GEOMETRIC ANALYSIS EXHIBITED NO ANOMALIES. NO CONCLUSION RELEVANT TO EXPERIENCE DETERMINED.

Description of Event or Problem · 1

FOLLOWING A LAPAROSCOPIC ANTI-REFLUX PROCEDURE, A PATIENT EXPERIENCED ONGOING GERD LEADING TO EXPLANT OF THE LINX DEVICE. THE LINX DEVICE WAS USED AS PART OF THE ANTI-REFLUX PROCEDURE. ANTI-REFLUX PROCEDURE AND LINX DEVICE IMPLANTATION OCCURRED WITHOUT ISSUE (B)(6) 2012. UNEVENTFUL DEVICE EXPLANT DUE TO ONGOING GERD ON (B)(6) 2016. DEVICE FOUND IN CORRECT POSITION/GEOMETRY. PATIENT UNDERWENT A FUNDOPLICATION IMMEDIATELY AFTER DEVICE EXPLANT.

Description of Event or Problem · 1

FOLLOWING A LAPAROSCOPIC ANTI-REFLUX PROCEDURE, A PATIENT EXPERIENCED ONGOING GERD LEADING TO EXPLANT OF THE LINX DEVICE. THE LINX DEVICE WAS USED AS PART OF THE ANTI-REFLUX PROCEDURE. ANTI-REFLUX PROCEDURE AND LINX DEVICE IMPLANTATION OCCURRED WITHOUT ISSUE (B)(6) 2012. UNEVENTFUL DEVICE EXPLANT DUE TO ONGOING GERD ON (B)(6) 2016. DEVICE FOUND IN CORRECT POSITION/GEOMETRY. PATIENT UNDERWENT A FUNDOPLICATION IMMEDIATELY AFTER DEVICE EXPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
665581 LINX REFLUX MANAGEMENT SYSTEM ANTI-REFLUX IMPLANT LEI TORAX MEDICAL, INC. LS14 3436 00855106005035

Patients

Seq Age Sex Outcome Treatment
1 58 YR Hospitalization| O