FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 6013766 · Received October 10, 2016

Report

Report Number
3004123209-2016-01514
Event Type
Malfunction
Date Received
October 10, 2016
Date of Event
August 26, 2016
Report Date
October 7, 2016
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE DEVICE DETAILED IN THIS REPORT WAS RETURNED TO HEARTSINE TECHNOLOGIES AS PART OF THE FSCA/RECALL Z-0124-2013 WITH NO ALLEGATION OF ANY FAULT WITH THE DEVICE. IT WAS THEREFORE INITIALLY ASSESSED AS NON-REPORTABLE HOWEVER SUBSEQUENT ENGINEERING INVESTIGATION REVEALED A FAULT WITH THE DEVICE WHICH WOULD CLASSIFY THE EVENT AS REPORTABLE. INVESTIGATION FOUND A FAULT WITH THE RETURNED DEVICE WHICH CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE FAULT WAS WITNESSED DURING THE INVESTIGATION. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE, WHICH IS WHY THE "UNKNOWN" BOX HAS BEEN CHECKED IN SECTION H8 OF THIS REPORT.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE WAS RETURNED TO HEARTSINE TECHNOLOGIES AS PART OF THE CURRENT FSCA/RECALL, FDA REFERENCE Z-0124-2013. NO FAULT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
665504 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM N/A

Patients

Seq Age Sex Outcome Treatment
1