SP6000
Report
- Report Number
- 9616031-2016-00005
- Event Type
- Injury
- Date Received
- October 10, 2016
- Date of Event
- September 12, 2016
- Report Date
- December 22, 2016
- Manufacturer
- GETINGE DISINFECTION AB
- Product Code
- FLH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING FILED UNDER EXEMPTION E2016015 BY THE MANUFACTURER GETINGE DISINFECTION AB, (REGISTRATION NO. 9616031) ON BEHALF OF THE IMPORTER GETINGE USA, INC., (REGISTRATION NO. 3004147784). ADDITIONAL INFORMATION WILL BE PROVIDED UPON RESULTS OF THE INVESTIGATION.
AN INVESTIGATION WAS PERFORMED ON THIS COMPLAINT. GETINGE RECEIVED A COMPLAINT WHERE IT WAS INDICATED THAT THE PATIENT RECEIVED CHEMICAL BURN ON HER POSTERIOR AFTER USING A BOWL PROCESSED IN WASHER DISINFECTOR MANUFACTURED BY THE GETINGE. THE PRODUCT INVOLVED IN THE INCIDENT IS GETINGE SP-6000 WASHER-DISINFECTOR. AFTER COMPLAINT REVIEW FOR THIS DEVICE WE ARE ABLE TO ESTABLISH THAT THIS ISSUE (CHEMICAL BURN FOR USER/PATIENT) CAN BE CONSIDERED TO BE A SINGLE, ISOLATED EVENT. (B)(4). PRIOR THE INCIDENT OCCURRED THE DEVICE WAS USED FOR DISINFECTING GOODS. AFTER THE PROCESS, THE GOODS WERE TAKEN OUT FROM THE WASHER AND USED BY THE PATIENT. THE MOST PROBABLE ROOT CAUSE OF THE CHEMICAL BURN OF THE PATIENT IS THE CUSTOMER (HOSPITAL) USING A WRONG DETERGENT. THE MATERIAL SAFETY DATA SHEET OF THE MENTIONED IN THE COMPLAINT DESCRIPTION GETINGE DETERGENT DOES NOT CONTAIN ANY POSSIBLY DANGEROUS INGREDIENTS. MOREOVER LOOKING AT THE DILUTION OF THE DETERGENT AFTER THE PROCESS IT IS CONSIDERED AS NOT POSSIBLE THAT OUR DETERGENT PLAYED A ROLE IN THE DESCRIBED EVENT. ADDITIONALLY THE CUSTOMER DID NOT ALLOW GETINGE REPRESENTATIVE TO INVESTIGATE AND EVALUATE THE DEVICE OR TO TAKE A SAMPLE AND INVESTIGATE THE DETERGENT, THEREFORE IT COULD BE INTERPRET AS POINTING TO A POTENTIAL USER ERROR. IT IS HIGHLY UNLIKELY THAT THE DETERGENT USED BY THE CUSTOMER IS THE ONE WHICH WAS MENTIONED AND RECOMMENDED BY THE GETINGE AND MOST PROBABLY FOR THIS REASON WE HAVE NOT RECEIVED ANY SAMPLES FOR AN EVALUATION. THE DEVICE WAS INSPECTED BY THE CUSTOMER SERVICE MANAGER AND AS IT WAS STATED NO MALFUNCTION WAS FOUND THAT WOULD HAVE PLAYED A ROLE IN CAUSING THE EVENT. TAKING UNDER CONSIDERATION ALL FACTORS, THE MOST PROBABLE ROOT CAUSE OF THE CHEMICAL BURN OF THE PATIENT IS CUSTOMER USING WRONG DETERGENT. IN SUMMARY, THE DEVICE WHICH PLAYED A ROLE IN THIS EVENT WAS USED FOR PATIENT AT THE TIME OF ISSUE OCCURRENCE, DUE TO DEVICE'S INTENDED USE. WE HAVE NOT RECEIVED ANY EVIDENCE THAT THE GETINGE DEVICE AND GETINGE DETERGENT FAILED TO MEET THEIR SPECIFICATION IN THE WAY THAT COULD BE DIRECTLY INVOLVED WITH THE OCCURRENCE OF THE REPORTED INCIDENT. THIS REPORT IS BEING FILED UNDER EXEMPTION E2016015 BY THE MANUFACTURER GETINGE DISINFECTION AB, (REGISTRATION NO. 9616031) ON BEHALF OF THE IMPORTER GETINGE USA, INC., (REGISTRATION NO. 3004147784).
IT WAS REPORTED BY THE GETINGE REPRESENTATIVE THAT THE PATIENT GOT CHEMICAL BURN AFTER USING THE BOWL WHICH WAS PREVIOUSLY DISINFECTED IN THE GETINGE'S DEVICE - TYPHOON. AFTER INCIDENT THE PATIENT WAS EXAMINED BY THE DOCTOR WHO CLASSIFIED THE INJURY AS CHEMICAL BURN. TECHNICAL SERVICE FROM THE FACILITY STATED THAT THE PRODUCT WHICH THEY PUT THE NIGHT BEFORE WAS GETINGE CLEAN FLUSHER RINSE LOT N° 11504.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 664734 | SP6000 | TYPHOON | FLH | GETINGE DISINFECTION AB | SP-6000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |