X SERIES
MDR report key: 6012162
·
Received October 7, 2016
Report
- Report Number
- 1220908-2016-02438
- Event Type
- Malfunction
- Date Received
- October 7, 2016
- Report Date
- September 19, 2016
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K112432
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL (B)(4) THE MALFUNCTION WAS DUPLICATED AND THE DEVICE'S PROCESSOR/BRIDGE/PACE BOARD WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE FAILED SELF-TEST FOR DEFIB AND PACER. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.