FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 6012162 · Received October 7, 2016

Report

Report Number
1220908-2016-02438
Event Type
Malfunction
Date Received
October 7, 2016
Report Date
September 19, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL (B)(4) THE MALFUNCTION WAS DUPLICATED AND THE DEVICE'S PROCESSOR/BRIDGE/PACE BOARD WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE FAILED SELF-TEST FOR DEFIB AND PACER. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663955 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES N/A

Patients

Seq Age Sex Outcome Treatment
1