FDA Adverse Event Summary report: N

APOLLO WHIRLPOOL BATH

MDR report key: 6012045 · Received October 7, 2016

Report

Report Number
2182947-2016-00001
Date Received
October 7, 2016
Date of Event
May 23, 2016
Report Date
June 16, 2016
Manufacturer
APOLLO CORPORATION
Product Code
ILJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THROUGH INVESTIGATING THE INCIDENT IT WAS DETERMINED THAT THE DEVICE WAS USED BY INDIVIDUALS WITHOUT PROPER TRAINING PRIOR TO THE EVENT. THE CHAIR WAS LIKELY REMOVED FROM THE REST OF THE DEVICE FOR CLEANING AND PLACED ON TOP OF THE LOWER SECTION OF THE CARRIER AS OPPOSED TO INTERFACING THE CHAIR'S BEARINGS WITH THE RAILS PROPERLY. DUE TO THIS THE CHAIR'S GUARDS LOCATED UNDER THE BEARINGS WERE RESTING ON THE RAILS INSTEAD OF UNDER THEM WHERE THEY WOULD HAVE WORKED WITH THE BEARINGS TO SECURE THE CHAIR IN PLACE. THE INCIDENT OCCURRED AS THE PATIENT SHIFTED TO THE SIDE AND THUS CAUSED THE CHAIR TO SHIFT AND FALL BETWEEN THE RAILS.

Description of Event or Problem · 1

PATIENT FELL FROM TRANSFER DEVICE CHAIR AND SUFFERED CONTUSION REQUIRING STITCHES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662897 APOLLO WHIRLPOOL BATH BATH CHAIR ILJ APOLLO CORPORATION 6000 N/A

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization