FDA Adverse Event
Summary report: N
APOLLO WHIRLPOOL BATH
MDR report key: 6012045
·
Received October 7, 2016
Report
- Report Number
- 2182947-2016-00001
- Date Received
- October 7, 2016
- Date of Event
- May 23, 2016
- Report Date
- June 16, 2016
- Manufacturer
- APOLLO CORPORATION
- Product Code
- ILJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THROUGH INVESTIGATING THE INCIDENT IT WAS DETERMINED THAT THE DEVICE WAS USED BY INDIVIDUALS WITHOUT PROPER TRAINING PRIOR TO THE EVENT. THE CHAIR WAS LIKELY REMOVED FROM THE REST OF THE DEVICE FOR CLEANING AND PLACED ON TOP OF THE LOWER SECTION OF THE CARRIER AS OPPOSED TO INTERFACING THE CHAIR'S BEARINGS WITH THE RAILS PROPERLY. DUE TO THIS THE CHAIR'S GUARDS LOCATED UNDER THE BEARINGS WERE RESTING ON THE RAILS INSTEAD OF UNDER THEM WHERE THEY WOULD HAVE WORKED WITH THE BEARINGS TO SECURE THE CHAIR IN PLACE. THE INCIDENT OCCURRED AS THE PATIENT SHIFTED TO THE SIDE AND THUS CAUSED THE CHAIR TO SHIFT AND FALL BETWEEN THE RAILS.
Description of Event or Problem · 1
PATIENT FELL FROM TRANSFER DEVICE CHAIR AND SUFFERED CONTUSION REQUIRING STITCHES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662897 | APOLLO WHIRLPOOL BATH | BATH CHAIR | ILJ | APOLLO CORPORATION | 6000 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |