PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-26683
- Event Type
- Malfunction
- Date Received
- October 7, 2016
- Date of Event
- April 11, 2014
- Report Date
- April 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS NOTED THAT THE PUMP WAS STUCK IN A MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY AND MOTOR POSITION ENCODER ERROR ALARM NOTED IN THE HISTORY FILE. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING THE TESTING. THE PUMP HAD CRACKED BATTERY TUBE THREADS AND SCRATCHED DISPLAY WINDOW. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA. (B)(4).
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM AND MOTOR POSITION ENCODER ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 112 MG/DL. THE DRIVE SUPPORT DISK WAS NORMAL AND THE PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELD. TROUBLESHOOTING WAS NOT ABLE TO RESOLVE THE ISSUE. THE DEVICE WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 664130 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |