FDA Adverse Event Death Summary report: N

TENDRIL STS

MDR report key: 6010962 · Received October 7, 2016

Report

Report Number
2938836-2016-11811
Event Type
Death
Date Received
October 7, 2016
Date of Event
November 26, 2015
Report Date
September 19, 2016
Manufacturer
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Product Code
NVN
PMA / PMN Number
P960013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPIRED. THE PATIENT DISAPPEARED WHILE WALKING AT THE RIVER SIDE AND HE WAS FOUND IN THE RIVER FOUR WEEKS LATER. IT WAS SUSPECTED THAT THE PATIENT HAD VF EPISODE AND PROBABLY FELLED INTO TO RIVER. REVIEW OF PROVIDED SESSION RECORD REVEALED, 10 VF, 1 VT-1, 14 AMS AND 40 NON-SUSTAINED RV OVERSENSING EPISODES SINCE THE PREVIOUS FOLLOW-UP, HOWEVER EGMS ARE AVAILABLE FOR THREE VF AND ONE VT-1 EPISODES. THE PATIENT HAD VT/VF EPISODES, POTENTIALLY AN ELECTRIC STORM. SINCE MANY OF THE VF EPISODES WITH SHOCK DELIVERY ARE OVERWRITTEN, CAN'T SEE IF THE DELIVERED SHOCKS WERE EFFECTIVE OR NOT. THE DEVICE SEEMS TO BEHAVE ACCORDING TO PROGRAMMED SETTINGS AND DESIGN OF ALGORITHMS. NO MALFUNCTION OF THE DEVICE IS SUSPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662758 TENDRIL STS PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) 2088TC/52 4183703

Patients

Seq Age Sex Outcome Treatment
1 Death 1458Q/86, (B)(4)| 7122Q/65, (B)(4)| CD3371-40Q, 1084842