FDA Adverse Event Malfunction Summary report: N

REVEAL XT

MDR report key: 6008498 · Received October 6, 2016

Report

Report Number
9614453-2016-06164
Event Type
Malfunction
Date Received
October 6, 2016
Date of Event
June 10, 2016
Report Date
July 14, 2016
Manufacturer
IPG MFG SWITZERLAND
Product Code
DSI
PMA / PMN Number
K103764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE ACTUAL LONGEVITY WAS LESS THAN 80% OF THE 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) COULD NOT BE INTERROGATED. THE ICM WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654574 REVEAL XT DETECTOR AND ALARM, ARRHYTHMIA DSI IPG MFG SWITZERLAND 9529

Patients

Seq Age Sex Outcome Treatment
1 84 YR