DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2016-72490
- Event Type
- Malfunction
- Date Received
- October 6, 2016
- Date of Event
- September 17, 2016
- Report Date
- September 17, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000019
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
DEXCOM WAS MADE AWARE ON (B)(6) 2016 , TO REPORT A MISSING SENSOR WIRE THAT OCCURRED ON (B)(6) 2016. THE SENSOR INSERTION WAS AT THE ABDOMEN ON THE (B)(6) 2016. THERE WAS NO REPORT ANY INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.
THREE SENSORS WERE RETURNED FOR EVALUATION; HOWEVER IT CANNOT BE DETERMINED WHICH IS THE DEVICE AT FAULT. A VISUAL INSPECTION WAS PERFORMED ON SENSOR (SERIAL# (B)(4)/ LOT# 5211761) AND FOUND THAT THE WIRE IS MISSING FROM THE SENSOR POD AND HOUSING PUCK. THE CUSTOMER'S COMPLAINT OF A MISSING/DETACHED SENSOR WIRE WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED. A VISUAL INSPECTION WAS PERFORMED ON SENSOR (SERIAL# (B)(4)/ LOT# 5211761) AND FOUND THAT THE WIRE IS MISSING FROM THE SENSOR POD AND HOUSING PUCK. THE CUSTOMER'S COMPLAINT OF A MISSING/DETACHED SENSOR WIRE WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED. A VISUAL INSPECTION WAS PERFORMED ON SENSOR (SERIAL# (B)(4)/ LOT# 5211761) AND FOUND THAT THE WIRE IS MISSING FROM THE SENSOR POD AND HOUSING PUCK. THE CUSTOMER'S COMPLAINT OF A MISSING/DETACHED SENSOR WIRE WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 658471 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | MDS | DEXCOM, INC. | 9500-27 | 5211761 | 00386270000019 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |