FDA Adverse Event Malfunction Summary report: N

CAPIOX RX25 OXYGENATOR EAST

MDR report key: 6004662 · Received October 6, 2016

Report

Report Number
1124841-2016-00350
Event Type
Malfunction
Date Received
October 6, 2016
Date of Event
September 14, 2016
Report Date
November 18, 2016
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
Product Code
DTZ
PMA / PMN Number
K130333
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TERUMO HAS RECEIVED THE DEVICE FOR EVALUATION; HOWEVER, THE INVESTIGATION HAS YET TO BE COMPLETED. TERUMO PLANS ON SUBMITTING A FOLLOW-UP REPORT WHEN THE INVESTIGATION IS COMPLETE AND WHEN MORE INFORMATION BECOMES AVAILABLE. (B)(4). CONCLUSION NOT YET AVAILABLE-EVALUATION IN PROGRESS.

Additional Manufacturer Narrative · 1

THE SAMPLE WAS RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO MANUFACTURING ANOMALIES. EVALUATION OF THE RETURNED SAMPLE CONFIRMED THAT THE LUER WAS NOT ABLE TO LOCK AROUND THE THREADS ON THE RESERVOIR PORT. WHEN THE DESIGN OF THE 3CX RESERVOIR LID WAS COMPARED AGAINST THE DESIGN OF THE (B)(6) PRODUCT, OR 1CX RESERVOIR, IT WAS FOUND THAT THE WALL THICKNESS OF THE PORTS DIFFERED. THE WALL THICKNESS OF THE WALL OPPOSITE OF THE THREADS ON THE PORTS IS THICKER ON THE (B)(6) RESERVOIR PORTS. THIS THICKER WALL HELPS KEEP THE THREADS OF THE PORT AND LUER ENGAGED. THE LUERED PORTS ON THE RESERVOIR LID ARE CHECKED IN PROCESS FOR ISO 594-1, THE INTERNAL SEALING TAPER, USING A CALIBRATED LUER GAUGE. ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW UP.

Description of Event or Problem · 1

THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING PRIME, IT WAS NOTED THAT THE MAQUET HEMOCONCENTRATOR LUER DOES NOT LOCK FIRMLY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
657736 CAPIOX RX25 OXYGENATOR EAST BLOOD GAS OXYGENATOR DTZ TERUMO CARDIOVASCULAR SYSTEMS CORPORATION 3CX*RX25RE TN30

Patients

Seq Age Sex Outcome Treatment
1