OT VERIO2 METER
Report
- Report Number
- 3008382007-2016-51707
- Event Type
- Malfunction
- Date Received
- October 6, 2016
- Report Date
- September 14, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
THE PRODUCT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON SEPTEMBER 14, 2016, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4) ALLEGING THE ONETOUCH VERIO 2 METER READ INACCURATELY ERRATIC. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. IT IS NOT SPECIFIED WHEN THE ALLEGED ISSUE BEGAN. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿230, 140, 300 AND 123MG/DL¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR PRECISION TESTING. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE WAS USED FOR TESTING. THE PATIENT CONFIRMED THE TEST STRIP VIAL WAS CRACKED OR BROKEN; HOWEVER, THERE WAS NO INDICATION OF MISUSE. BASED ON THE INFORMATION PROVIDED, THERE IS NO INDICATION THAT THE REPORTED ISSUE CAUSED OR CONTRIBUTED TO A SERIOUS INJURY SINCE THE PATIENT DENIED DEVELOPING SYMPTOMS OR RECEIVING MEDICAL TREATMENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED INACCURACY REMAINED UNRESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 655144 | OT VERIO2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 4005817 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |