THERMOCOOL® SMARTTOUCH® BI-DIRECTIONAL NAVIGATION CATHETER
Report
- Report Number
- 9673241-2016-00657
- Event Type
- Death
- Date Received
- October 5, 2016
- Date of Event
- September 9, 2016
- Report Date
- September 19, 2016
- Manufacturer
- BIOSENSE WEBSTER, INC. (JUAREZ)
- Product Code
- LPB
- PMA / PMN Number
- P030031/S053
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS DISCARDED, THEREFORE NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED AND DEVICE MALFUNCTION CANNOT BE CONFIRMED. SINCE NO LOT NUMBER WAS PROVIDED, NO DEVICE HISTORY RECORD (DHR) REVIEW COULD BE PERFORMED. CONCOMITANT PRODUCTS: CARTO 3 SYSTEM. CIRCA SCIENTIFIC MULTIPOLAR TEMPERATURE PROBE. FULL UDI # INFORMATION UNAVAILABLE SINCE THE LOT NUMBER IS UNKNOWN. (B)(4). ARE RELATED TO THE SAME INCIDENT.
IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT UNDERWENT AN ABLATION PROCEDURE FOR ATRIAL FIBRILLATION WITH A THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHETER AND SUFFERED AN ESOPHAGEAL FISTULA RESULTING IN THE PATIENT'S DEATH. POST-PROCEDURE, AN ESOPHAGEAL FISTULA WAS DIAGNOSED. THE DEATH TOOK PLACE AN UNKNOWN AMOUNT OF TIME AFTER. THERE IS NO INFORMATION REGARDING THE DETAILED COURSE OF EVENTS OR AN AUTOPSY. PHYSICIAN'S OPINION REGARDING THE CAUSE OF THE ADVERSE EVENT IS THAT IT WAS PROCEDURE-RELATED AND NOT THE RESULT OF A PRODUCT MALFUNCTION. PHYSICIAN IS NOT CERTAIN WHICH ABLATION SITE WAS THE SITE OF INJURY. GENERATOR WAS IN POWER CONTROL MODE AT 35 WATTS ON THE ANTERIOR WALL AND 20-25 WATTS ON THE POSTERIOR WALL. TEMPERATURE WARNING WAS SET AT 45 DEGREES CELSIUS AND CUT-OFF AT 50 DEGREES CELSIUS. ABLATION TIME AT THE SITE OF INJURY WAS NOT REPORTED, AS THERE WERE NO PROLONGED ABLATION SITES ASSOCIATED WITH INCREASED TEMPERATURES. CIRCA SCIENTIFIC MULTIPOLAR TEMPERATURE PROBE WAS USED TO PREVENT ESOPHAGEAL INJURY. ATTEMPTS WERE MADE TO AVOID THE ESOPHAGUS WHILE COMPLETING A POSTERIOR BOX. SPI VALUE IS UNKNOWN. THE THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHETER WAS NOT IN CLOSE PROXIMITY TO ANOTHER CATHETER. THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHETER WAS ZEROED AFTER THE INITIAL WARM-UP PHASE, POST-CONNECTION TO THE CARTO 3 PIU. CARTO 3 SYSTEM DID NOT INDICATE TO RE-ZERO THE CATHETER. THERE WERE NO ERRORS REPORTED ON ANY BWI EQUIPMENT DURING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 652461 | THERMOCOOL® SMARTTOUCH® BI-DIRECTIONAL NAVIGATION CATHETER | CARDIAC ABLATION PERCUTANEOUS CATHETER | LPB | BIOSENSE WEBSTER, INC. (JUAREZ) | D-1327-05-S | UNK_D-1327-05-S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Death |