TRUEMETRIX
Report
- Report Number
- 1000113657-2016-01678
- Event Type
- Malfunction
- Date Received
- October 5, 2016
- Date of Event
- September 12, 2016
- Report Date
- October 14, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE IS UNDERGOING AN EVALUATION BUT HAS NOT YET BEEN COMPLETED.
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECT FOUND ON RETURNED METER. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER'S TEST STRIPS HAD A POOR STORAGE(KITCHEN).
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 TO 125 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/30/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). CUSTOMER HAS NOT HAD ANY RECENT CHANGES IN DIET, EXERCISE, OR MEDICATIONS. CUSTOMER TAKES JANUVIA FOR DIABETES MANAGEMENT.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 TO 125MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/30/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). CUSTOMER HAS NOT HAD ANY RECENT CHANGES IN DIET, EXERCISE, OR MEDICATIONS. CUSTOMER TAKES JANUVIA FOR DIABETES MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 652363 | TRUEMETRIX | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUEMETRIX | MT1699 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |