MODULAR EBS KNEE JOINT
Report
- Report Number
- 9616494-2016-00025
- Event Type
- Malfunction
- Date Received
- October 5, 2016
- Report Date
- September 21, 2016
- Manufacturer
- OTTO BOCK HEALTHCARE GMBH
- Product Code
- ISY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE EVALUATION WAS CARRIED OUT AT THE MANUFACTURER'S SERVICE CENTER ON SEPTEMBER 26, 2016. SEVERAL BOLTS ARE USED TO FIX THE DIFFERENT PARTS OF THE KNEE JOINT WHEN IT IS MOUNTED. THE THREADS OF THE BOLTS ARE COVERED WITH AN TWO-COMPONENT-ADHESIVE, WHICH IS HARDENING ONCE THE BOLT IS TIGHTENED. THIS IS TO ASSURE THAT THE PARTS ARE FIXED PERMANENTLY. AFTER EVALUATION WE FOUND THAT THE HEXAGON SOCKETS OF THE BOLTS ARE DAMAGED. WE ASSUME THAT THE BOLTS WERE TIGHTENED OR LOOSENED WITH NOT SUFFICIENT TOOLS SO THAT THE DAMAGE COULD OCCUR. BOTH BOLTS IN THE UPPER POSTERIOR PART OF THE KNEE JOINT WERE LOOSE. DUE TO FURTHER USAGE OF THE KNEE JOINT WITH LOOSE BOLTS THE ANTERIOR FRAME WAS BENT. NO FALL OR INJURY OCCURRED DUE TO THIS FAILURE, BUT IT COULD HAVE LED TO PATIENT INJURY.
KNEE JOINT WAS SENT IN FOR REPAIR. CUSTOMER STATED THAT THE JOINT IS NOT FUNCTIONING PROPERLY. NO FALL OR SIGNIFICANT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 652358 | MODULAR EBS KNEE JOINT | MODULAR EBS KNEE JOINT | ISY | OTTO BOCK HEALTHCARE GMBH | 3R60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |