VERSACARE FRAME
Report
- Report Number
- 1824206-2016-00431
- Event Type
- Malfunction
- Date Received
- October 4, 2016
- Date of Event
- September 16, 2016
- Report Date
- September 16, 2016
- Manufacturer
- HILL-ROM BATESVILLE
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICIAN FOUND THE EXTERNAL ALARM NEEDED TO BE REPLACED. PER THE HILL-ROM SERVICE MANUAL, PERFORM ANNUAL PREVENTIVE MAINTENANCE PROCEDURES TO MAKE SURE ALL VERSACARE® BED COMPONENTS ARE FUNCTIONING AS ORIGINALLY DESIGNED. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. EXAMINE THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE ARE SET. REPLACE AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2016. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE EXTERNAL ALARM TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKE NOT SET ALARM IS NOT SOUNDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 648701 | VERSACARE FRAME | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM BATESVILLE | P3200K000453 |