FDA Adverse Event Malfunction Summary report: N

REFOBACIN BONE CEMENT R 40X1

MDR report key: 5995967 · Received October 4, 2016

Report

Report Number
3006946279-2016-00354
Event Type
Malfunction
Date Received
October 4, 2016
Date of Event
August 30, 2016
Report Date
March 30, 2017
Manufacturer
BIOMET FRANCE S.A.R.L.
Product Code
MBB
PMA / PMN Number
PK150850
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS USER FACILITY IS OUTSIDE OF THE UNITED STATES. THE NECESSARY MANUFACTURING HISTORY TO REVIEW WAS NOT PROVIDED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. ZIMMER BIOMET FRANCE AND PACKAGING SUPPLIER ARE IN THE PROCESS OF INITIATING MODIFICATIONS TO ADDRESS REPORTED ISSUE. THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME PATIENT (REFERENCE 3006946279-2016-00354 / 00355).

Additional Manufacturer Narrative · 1

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. (B)(4). COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. UPON EXAMINATION OF THE PRODUCT, IT WAS NOTED THAT A PORTION OF THE INNER PACKAGING WAS NOT SEALED. THE OUTER POUCH WAS NOT DAMAGED. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. INVESTIGATION RESULTS CONCLUDED THAT THE REPORTED EVENT WAS DUE TO AN INADEQUATE SEAL FORMED WITH SEALING MATERIAL USED. A SUMMARY OF THE INVESTIGATION HAS BEEN SENT TO THE COMPLAINANT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTRIBUTE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN BOX OF BONE CEMENT WAS OPENED, THE INNER PACKAGE OF POWDER WAS LEAKING INTO THE OUTER PACKAGING. THERE WAS NO PATIENT INJURY AND A TEN MINUTE DELAY IN A PROCEDURE AS A RESULT OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
648449 REFOBACIN BONE CEMENT R 40X1 BONE CEMENT, ANTIBIOTIC MBB BIOMET FRANCE S.A.R.L. N/A A418AF2405

Patients

Seq Age Sex Outcome Treatment
1