PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-23048
- Event Type
- Malfunction
- Date Received
- October 3, 2016
- Date of Event
- October 18, 2013
- Report Date
- October 18, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED ON MOTOR HOME SWITCH. NO MOISTURE DAMAGE WAS FOUND ON ELECTRONIC ASSY. THE PUMP WAS UNABLE TO VERIFY THE MOTOR POSITION ENCODER ERROR ALARM DUE TO MOTOR ERROR. THE PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA. (B)(4).
THE CUSTOMER REPORTED VIA PHONE CALL INDICATING MOTOR ERROR ALARM AND MOTOR POSITION ENCODER ERROR FROM THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 214 MG/DL. THE CUSTOMER STATED THAT THE PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELD. THE CUSTOMER WAS ADVISED TO CHECK IF THE DRIVE SUPPORT CAP IS PROTRUDED, LOOSE, STICKING OUT OR FLUSH. THE CUSTOMER REPORTED THE DRIVE SUPPORT CAP TO APPEAR NORMAL. THE CUSTOMER STATED THAT THERE WAS A MOTOR POSITION ENCODER ERROR IN THE ALARM HISTORY. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND TO REVERT TO A BACKUP PLAN PER HEALTH CARE PROVIDER'S INSTRUCTIONS. THE CUSTOMER WILL BE SENT A REPLACEMENT PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 644589 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |