FDA Adverse Event Malfunction Summary report: N

SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK

MDR report key: 5993547 · Received October 3, 2016

Report

Report Number
3003761017-2016-00339
Event Type
Malfunction
Date Received
October 3, 2016
Date of Event
September 20, 2016
Report Date
September 20, 2016
Manufacturer
SYNCARDIA SYSTEMS, LLC
Product Code
LOZ
PMA / PMN Number
P030011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS ALLEGED FAILURE MODE POSES A LOW RISK TO THE PATIENT SINCE THE DRAINED BATTERY DID NOT PREVENT THE COMPANION 2 DRIVER FROM PERFORMING ITS LIFE-SUSTAINING FUNCTIONS. THE COMPANION 2 DRIVER HAS A REDUNDANT, ALTERNATE POWER SOURCE OF EXTERNAL WALL POWER AND BACKUP EXTERNAL BATTERIES. THE COMPANION EXTERNAL BATTERY WILL BE RETURNED TO SYNCARDIA FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WILL BE PROVIDED IN A FOLLOW-UP MDR. (B)(4) (2 OF 2).

Additional Manufacturer Narrative · 1

THE COMPANION EXTERNAL BATTERY WAS RETURNED TO SYNCARDIA FOR EVALUATION. THE SYSTEM MANAGEMENT BUS (SMBUS) DATA WAS REVIEWED AND REVEALED NO ANOMALOUS BEHAVIOR OR VALUES. THE REPORTED LOW CAPACITY CONDITION COULD NOT BE DUPLICATED OR CONFIRMED. THE EXTERNAL BATTERY WAS TESTED, INCLUDING BATTERY DISCHARGE, AND EVALUATION DETERMINED THAT THE EXTERNAL BATTERY FUNCTIONED AS INTENDED WITH NO EVIDENCE OF A DEVICE MALFUNCTION. THIS ISSUE WILL CONTINUE TO BE MONITORED AND TRENDED AS PART OF THE CUSTOMER EXPERIENCE PROCESS. SYNCARDIA HAS COMPLETED ITS EVALUATION OF THIS COMPLAINT AND IS CLOSING THIS FILE. (B)(4) FOLLOW-UP REPORT 1 (2 OF 2).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE COMPANION EXTERNAL BATTERY LOST ITS CHARGE AFTER 15 MINUTES WHILE IN A COMPANION 2 DRIVER SUPPORTING A PATIENT. THE CUSTOMER ALSO REPORTED THAT A SUBSEQUENT TEST REVEALED THE COMPANION EXTERNAL BATTERY LOST ITS CHARGE AFTER 30 MINUTES. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE COMPANION EXTERNAL BATTERY LOST ITS CHARGE AFTER 15 MINUTES WHILE IN A COMPANION 2 DRIVER SUPPORTING A PATIENT. THE CUSTOMER ALSO REPORTED THAT A SUBSEQUENT TEST REVEALED THE COMPANION EXTERNAL BATTERY LOST ITS CHARGE AFTER 30 MINUTES. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
644635 SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK BATTERY PACK LOZ SYNCARDIA SYSTEMS, LLC

Patients

Seq Age Sex Outcome Treatment
1 37 YR