FDA Adverse Event Malfunction Summary report: N

ARCHITECT IPTH

MDR report key: 5993441 · Received October 3, 2016

Report

Report Number
3002809144-2016-00083
Event Type
Malfunction
Date Received
October 3, 2016
Date of Event
June 15, 2016
Report Date
April 24, 2017
Manufacturer
ABBOTT GERMANY
Product Code
CEW
PMA / PMN Number
K063232
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER STATED THAT THE ARCHITECT INTACT PTH ASSAY RESULTS WERE ELEVATED WHEN TESTING PATIENT RESULTS. A REVIEW OF COMPLAINT REPORTS RECEIVED TO DATE IDENTIFIED NORMAL COMPLAINT ACTIVITY FOR THIS ISSUE. ACCURACY TESTING WAS PERFORMED WITH A FILE KIT OF LOT 00715G000 AND ALL RESULTS MET ACCEPTANCE CRITERIA. A REVIEW OF LABELING FOUND THAT THE CUSTOMER ISSUE OF FALSELY ELEVATED IPTH PATIENT RESULTS WAS ADEQUATELY ADDRESSED. THERE WAS NO PRODUCT DEFICIENCY IDENTIFIED FOR THIS ISSUE. IN ADDITION, THERE IS NOT ENOUGH INFORMATION TO REASONABLY SUGGEST A MALFUNCTION HAD OCCURRED. BASED ON THIS INVESTIGATION IT HAS BEEN DETERMINED THAT THE ARCHITECT INTACT PTH ASSAY, LOT 00715G000 IS PERFORMING AS INTENDED.

Additional Manufacturer Narrative · 1

THE DEVICE EVALUATION WAS REASSESSED AND CONCLUDED THAT A MALFUNCTION OCCURRED, THEREFORE THE DEVICE WAS NOT PERFORMING AS INTENDED, AND CONCLUSIONS CODE WAS CORRECTED FROM (B)(4).

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8K25-25 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 8K25-27.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE ARCHITECT ANALYZER GENERATED FALSELY ELEVATED IPTH RESULTS ON ONE PATIENT. THE RESULTS PROVIDED WERE: 15JUN2016 = 107PG/ML / 06JUL2016 = 116PG/ML / RESULTS PRIOR TO LUMPECTOMY ON 12SEP2016 = 60.0PG/ML / ANOTHER ANALYZER = 58.8PG/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
645997 ARCHITECT IPTH INTACT PARATHYROID HORMONE CEW ABBOTT GERMANY 00715G000

Patients

Seq Age Sex Outcome Treatment
1 41 YR ARCHITECT I2000SR ANALYZER, LIST # 03M74-01,| SERIAL # (B)(4)