PURELY YOURS ULTRA
Report
- Report Number
- 3009974348-2016-00224
- Event Type
- Injury
- Date Received
- September 29, 2016
- Date of Event
- August 25, 2016
- Report Date
- September 6, 2016
- Manufacturer
- AMEDA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K973501
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE RETURNED PRODUCT WAS ASSESSED FOR INDICATIONS OF MALFUNCTION OR THERMAL EVENT. THE RETURNED PRODUCT WAS ASSESSED FOR FUNCTIONALITY AND MET FUNCTIONAL SPECIFICATIONS. BATTERIES WERE NOT RETURNED WITH THE PUMP. ADDITIONAL TESTING CONFIRMED THAT BATTERIES MAY GET HOT AND LEAK WHEN THE UPPER MIDDLE BATTERY IS PLACED INCORRECTLY, WITH THE POSITIVE AND NEGATIVE END REVERSED.
CUSTOMER CONTACTED AMEDA, INC. TO REPORT SUSTAINING A BURN TO THE PAD OF HER LEFT RING FINGER WHEN A HOT BATTERY POPPED OUT OF THE BATTERY COMPARTMENT OF HER PURELY YOURS ULTRA BREAST PUMP AFTER SENSING THEY WERE BECOMING OVERHEATED. THIS OCCURRED WHILE MOM WAS PUMPING IN HER CAR ON (B)(6) 2016. CUSTOMER STATES SHE STRUGGLED WITH THE BATTERIES REMAINING SECURELY INSIDE THE BATTERY COMPARTMENT IN ORDER TO CLOSE THE COMPARTMENT DOOR. CUSTOMER REPORTS THE BURN AS MINOR, WASHED FINGER WITH CLEAN, COOL WATER AND DID NOT SEEK OUT MEDICAL ATTENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 640113 | PURELY YOURS ULTRA | ELECTRIC BREAST PUMP | HGX | AMEDA, INC. | 24501879 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |