O-ARM 1000 IMAGING SYSTEM
Report
- Report Number
- 1723170-2016-04198
- Event Type
- Malfunction
- Date Received
- September 28, 2016
- Date of Event
- August 24, 2015
- Report Date
- September 28, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. BENCH LEVEL TESTING FOUND THAT THE LEMO CONNECTOR PIN 5 BROWN WIRE TO THE YELLOW SIDE CONNECTOR PIN 3 WAS DISCONNECTED FROM THE LEMO HOUSING RESULTING IN AN OPEN. THE CABLE WAS REPLACED AND THE DEVICE PASSED SYSTEM CHECKOUT TESTS. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(4) DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
A MEDTRONIC REPRESENTATIVE REPORTED THAT THE IMAGING SYSTEM'S UMBILICAL CABLE HAD A LOOSE CONNECTION. AT SOME TIMES IT WOULD NOT CONNECT PROPERLY. THE CABLE SHOWED VISIBLE SIGNS OF STRESS. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 635622 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |