FDA Adverse Event Malfunction Summary report: N

INFINITY ACUTE CARE SYSTEM (M540)

MDR report key: 5984627 · Received September 28, 2016

Report

Report Number
1220063-2016-00030
Event Type
Malfunction
Date Received
September 28, 2016
Date of Event
August 22, 2016
Report Date
November 2, 2016
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC
Product Code
CBK
PMA / PMN Number
K113798
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE IACS LOGS FROM THE EVENT WERE REVIEWED BY ENGINEERING. A VIDEO CAPTURING THE ISSUE WAS PROVIDED. AFTER REVIEWING THE PROVIDED VIDEO AND CORRESPONDING LOGS, IT COULD BE CONFIRMED THAT THE PULSE TONES AND AUDIBLE ALARMS WERE NOT WORKING DURING THE EVENT AFTER BEING IN STANDBY. THE NO AUDIO ISSUE AT BOTH THE COCKPIT AND M540 WAS REPRODUCED DURING THE INVESTIGATION. IT HAS BEEN DETERMINED THAT THERE IS A BUG IN THE MICROSOFT WINDOWS AUDIO SERVICE SUBSYSTEM THAT CAUSES A LOSS OF ALL AUDIO. A FIELD SAFETY CORRECTIVE ACTION HAS BEEN RELEASED TO CURRENT USERS OF IACS VG5 FOR A MANDATORY DOWNGRADE TO VG4. THE FIELD SAFETY CORRECTIVE ACTION RELEASED FOR THIS ISSUE IS NOT APPLICABLE FOR THE DEVICES IN THE US MARKET WHERE THE AFFECTED SOFTWARE VERSION IS NOT USED.

Additional Manufacturer Narrative · 1

THE INVESTIGATION WAS STARTED BUT HAS NOT YET BEEN CONCLUDED. THE INVESTIGATION RESULT WILL BE REPORTED IN THE FOLLOW UP REPORT.

Description of Event or Problem · 1

PLEASE REFER TO THE INITIAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT, AFTER INSTALLATION 9:30PM (B)(6), ALL FUNCTIONS OF IACS ARE FINE, HOWEVER, THE NEXT MORNING THE CLINICAL SPECIALIST NOTICED THAT WHILST THE M540 WAS ACOUSTICALLY ALARMING (WHICH IT SHOULDN'T IN AN IACS CONFIGURATION), AND THE COCKPIT WAS NOT ANNUNCIATING AN ALARM. FURTHER COMMUNICATION WITH CUSTOMER INDICATED THE M540 INTERMITTENTLY HAD NO AUDIBLE ALARM. THERE IS NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
634650 INFINITY ACUTE CARE SYSTEM (M540) PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT CBK DRAEGER MEDICAL SYSTEMS, INC NA NA

Patients

Seq Age Sex Outcome Treatment
1