FDA Adverse Event Malfunction Summary report: N

MUSTANG¿

MDR report key: 5984152 · Received September 28, 2016

Report

Report Number
2134265-2016-08536
Event Type
Malfunction
Date Received
September 28, 2016
Date of Event
September 5, 2016
Report Date
September 5, 2016
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
FGE
PMA / PMN Number
K110122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA) UTILIZING A CONTRALATERAL APPROACH. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SFA. AFTER PRE-DILATATION WAS PERFORMED WITH 2MM COYOTE BALLOON CATHETER, A 5.0X100,135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON THE FIRST INFLATION AT 20 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A 4MM COYOTE BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
635978 MUSTANG¿ CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - GALWAY H74939171051010 18705300

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE: AGOSAL| INFLATION DEVICE: GOODMAN| INTRODUCER SHEATH: DESTINATION