FDA Adverse Event Malfunction Summary report: N

SURGIWAND*II 5MM SUC/IRR W/CAUT/SPATULA

MDR report key: 5983109 · Received September 28, 2016

Report

Report Number
2647580-2016-00766
Event Type
Malfunction
Date Received
September 28, 2016
Date of Event
August 31, 2016
Report Date
September 2, 2016
Manufacturer
COVIDIEN, FORMERLY USSC PUERTO RICO INC
Product Code
GEI
PMA / PMN Number
K961771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED ONE DEVICE BY THE ACCOUNT. THIS EVALUATION WAS BASED ON A TECHNICAL REVIEW OF ALL DATA RECEIVED FROM THE SITE, A PMV REVIEW OF MANUFACTURING RECORDS, A PMV REVIEW OF COMPLAINT TRENDS, AND AN EVALUATION OF THE RETURNED DEVICE. THE VISUAL INSPECTION OF THE DEVICE NOTED THAT THE COLLAR WAS DAMAGED ON THE INSTRUMENT; THE COLLAR WAS ATTACHED TO THE DEVICE BUT A PLASTIC TAB WAS BROKEN. DURING A FUNCTIONAL EVALUATION, THE INSTRUMENT WAS TESTED FOR ELECTRICAL CONTINUITY AND PROPER SUCTION AND IRRIGATION FUNCTION WITH ACCEPTABLE RESULTS. THE INSTRUMENT ALSO PASSED TWO AIR LEAK TESTS. REPLICATION OF THE OBSERVED DAMAGE MAY OCCUR IF EXCESS FORCE IS USED WHEN THE COLLAR IS PUSHED FORWARD TO COVER THE L-HOOK. BASED ON THE PRODUCT ANALYSIS, THE FAILURE WAS CONFIRMED TO BE ATTRIBUTED TO THE REPORTED EVENT. SHOULD NEW INFORMATION BECOME AVAILABLE, THE FILE WILL BE RE-OPENED AND THE INVESTIGATION SUMMARY WILL BE AMENDED AS APPROPRIATE.

Description of Event or Problem · 1

ACCORDING TO THE REPORTER, DURING PREPARATION FOR A PROCEDURE, THE WHITE PART WAS DISENGAGED FROM THE CONNECTION OF THE SHEATH OUT-OF-PACKAGE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE AND THERE WAS NO INJURY TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
635862 SURGIWAND*II 5MM SUC/IRR W/CAUT/SPATULA ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI COVIDIEN, FORMERLY USSC PUERTO RICO INC 178099 P6E0318X

Patients

Seq Age Sex Outcome Treatment
1