FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 5983064 · Received September 28, 2016

Report

Report Number
2531779-2016-27034
Event Type
Malfunction
Date Received
September 28, 2016
Report Date
September 6, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 11/09/2016 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/20/2016 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP WAS POWERED ON AND EMITTED A CS 069 CALL SERVICE ALARM IMMEDIATELY UPON ATTEMPTING A REWIND STEP. THE CALL SERVICE ALARM WAS RECORDED IN THE BLACK BOX DATA AS A CS 087 FAILURE. INVESTIGATION REVEALED THAT A LANGUAGE CORRUPTION OCCURRED AT A COMPONENT ON THE PRINTED CIRCUIT BOARD RESULTING IN A CALL SERVICE ALARM. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY CAP WAS RETURNED STUCK TO THE PUMP AND HAD TO BE FORCIBLY REMOVED. THE BATTERY CAP COIN SLOT WAS FOUND TO BE DAMAGED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
636156 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR