ANIMAS VIBE
Report
- Report Number
- 2531779-2016-27034
- Event Type
- Malfunction
- Date Received
- September 28, 2016
- Report Date
- September 6, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 11/09/2016 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/20/2016 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP WAS POWERED ON AND EMITTED A CS 069 CALL SERVICE ALARM IMMEDIATELY UPON ATTEMPTING A REWIND STEP. THE CALL SERVICE ALARM WAS RECORDED IN THE BLACK BOX DATA AS A CS 087 FAILURE. INVESTIGATION REVEALED THAT A LANGUAGE CORRUPTION OCCURRED AT A COMPONENT ON THE PRINTED CIRCUIT BOARD RESULTING IN A CALL SERVICE ALARM. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY CAP WAS RETURNED STUCK TO THE PUMP AND HAD TO BE FORCIBLY REMOVED. THE BATTERY CAP COIN SLOT WAS FOUND TO BE DAMAGED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 636156 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |