FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 5982367 · Received September 28, 2016

Report

Report Number
9611451-2016-00599
Event Type
Malfunction
Date Received
September 28, 2016
Date of Event
August 23, 2016
Report Date
September 1, 2016
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED VERTICAL CRACKS AROUND THE BASE OF THE CHAMBER DOME. ONE CRACK WAS BELOW ONE OF THE PORTS AND THE OTHER ONE WAS AT THE HINGE BRACKET. WHITE RESIDUE WAS PRESENT AT DIFFERENT LOCATIONS ON THE INNER AND OUTER SURFACE OF THE CHAMBER DOME. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT 160128. CONCLUSION: THE WHITE RESIDUE SUGGESTS THAT THE DAMAGE WAS CAUSED BY THE CHAMBER COMING INTO CONTACT WITH A SOLUTION CONTAINING ETHANOL/ALCOHOL WHICH HAS RESULTED IN ENVIRONMENTAL STRESS CRACKING OF THE CHAMBER DOME. EVERY MR290 CHAMBER IS PRESSURE TESTED TO 200 CMH2O FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: DO NOT SOAK, WASH, STERILIZE, OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS, OR HAND SANITIZERS. SET APPROPRIATE VENTILATOR ALARMS. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN MR290V HUMIDIFICATION CHAMBER WAS FOUND TO BE CRACKED AFTER 14 DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
637177 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 2100021582

Patients

Seq Age Sex Outcome Treatment
1