SORIN HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2016-00677
- Date Received
- September 27, 2016
- Report Date
- March 30, 2017
- Manufacturer
- SORIN GROUP DEUTSCHLAND
- Product Code
- DWC
- PMA / PMN Number
- K052601
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE FACILITY DID NOT PROVIDE THE EVENT DATE. THE INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF AND WHEN MADE AVAILABLE. SORIN IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF SORIN HEATER-COOLER DEVICES. THE Z NUMBER IS (B)(4). SORIN GROUP (B)(4) MANUFACTURES THE SORIN HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE SORIN HEATER-COOLER SYSTEM 3T TESTED POSITIVE FOR NONTUBERCULOUS MYCOBACTERIUM (NTM). THERE IS NO KNOWN PATIENT INVOLVEMENT. THE CUSTOMER ALSO STATED IN THE REPORT THAT THREE CONSECUTIVE DISINFECTION CYCLES WERE PERFORMED USING MINNCARE DUE TO THE CONTAMINATION. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.
(B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). DURING FOLLOW-UP COMMUNICATION, THE CUSTOMER INFORMED LIVANOVA (B)(4) THAT THE HOSPITAL MANAGEMENT WAS CONCERNED WITH SHARING SOME OF THE REQUESTED INFORMATION WITH THE COMPANY. HOWEVER, IT WAS DISCLOSED THAT THE HEATER-COOLER 3T SYSTEMS AT THE FACILITY ARE NOT IN REGULAR USE AND ARE PUT IN A HOLDING PATTERN WHILE THE FACILITY WAITS FOR THE RESULTS TO COME BACK FROM THE LAB. HOWEVER, THE DEVICES MAY STILL BE USED IF IT IS REQUIRED DUE TO A LACK OF EQUIPMENT. NO FURTHER INFORMATION HAS BEEN PROVIDED. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. AS CORRECTIVE ACTION, A FIELD SAFETY NOTICE HAS BEEN RELEASED TO REMIND OUR CUSTOMERS ABOUT THE IMPORTANCE OF ADHERING TO THE WATER MANAGEMENT AND DISINFECTION PROCEDURE OUTLINED IN THE CURRENT INSTRUCTIONS FOR USE (IFU). DEVICES NOT RETURNED.
SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE SORIN HEATER-COOLER SYSTEM 3T TESTED POSITIVE FOR NONTUBERCULOUS MYCOBACTERIUM (NTM). THERE IS NO KNOWN PATIENT INVOLVEMENT.