PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 5981691 · Received September 27, 2016

Report

Report Number
2032227-2016-24312
Event Type
Malfunction
Date Received
September 27, 2016
Date of Event
September 6, 2016
Report Date
September 6, 2016
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED DISPLACEMENT TEST FOLLOWED BY A MOTOR ERROR ALARM DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. THE PUMP WAS UNABLE TO PERFORM FUNCTIONAL TEST INCLUDING SELF-TEST, UNEXPECTED ERROR TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT TEST OR VERIFY ALARM DUE TO MOTOR ERROR ALARMS. THE PUMP WAS RECEIVED WITH MISSING END CAP STICKER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL OF SPI TO MOTOR PIC COMMUNICATION ERROR, EXCESSIVE NO DELIVERIES AND MRI EXPOSURE FROM THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 86 MG/DL. THE CUSTOMER STATED THAT SHE WENT IN FOR AN MRI AND WORE IT IN THE MACHINE. THE CUSTOMER STATED THAT THE PUMP GAVE REPEATED ALARMS. THE CUSTOMER WILL RETURN THE PUMP FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631881 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 25 YR