PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 5981372
·
Received September 27, 2016
Report
- Report Number
- 2032227-2016-21965
- Event Type
- Malfunction
- Date Received
- September 27, 2016
- Date of Event
- November 14, 2013
- Report Date
- November 14, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS NOTED THAT THE PUMP ALARMED MOTOR ERROR DURING REWIND AND MOTOR POSITION ENCODER ERROR CONFIRMED IN THE HISTORY FILE DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. THEY WERE UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO THE MOTOR ERROR ALARM. THE PUMP HAD SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE PUMP HAD MOTOR ERROR ALARM AND MOTOR POSITION ENCODER ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 103 MG/DL. TROUBLESHOOTING WAS NOT ABLE TO RESOLVE THE ISSUE. THE DEVICE WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632747 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |