FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 5981372 · Received September 27, 2016

Report

Report Number
2032227-2016-21965
Event Type
Malfunction
Date Received
September 27, 2016
Date of Event
November 14, 2013
Report Date
November 14, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS NOTED THAT THE PUMP ALARMED MOTOR ERROR DURING REWIND AND MOTOR POSITION ENCODER ERROR CONFIRMED IN THE HISTORY FILE DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. THEY WERE UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO THE MOTOR ERROR ALARM. THE PUMP HAD SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE PUMP HAD MOTOR ERROR ALARM AND MOTOR POSITION ENCODER ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 103 MG/DL. TROUBLESHOOTING WAS NOT ABLE TO RESOLVE THE ISSUE. THE DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632747 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 37 YR