ALLURA XPER FD10

MDR report key: 597986 · Received May 3, 2005

Report

Report Number
1217116-2005-00009
Event Type
Other
Date Received
May 3, 2005
Date of Event
March 21, 2005
Report Date
April 29, 2005
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
IZI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

"EQUIPMENT BEGAN TO DISPLAY AN ALARM, "FLAT DETECTOR OVERHEATING, FINISH CASE, CALL SERVICE" (A SYSTEM WARNING MESSAGE). CATH LAB TEAM LEADER WAS NOTIFIED, WHO PLACED CALL TO PHILIPS MEDICAL SYSTEMS. TEAM LEAD TROUBLESHOT ALARM WITH RTAC SUPPORT TO DETERMINE REASON FOR ALARM. COOLANT LOW? NO, GAUGE READING INDICATED THAT THE TEMP. LEVEL WAS CLIMBING ON THE FLAT DETECTOR. TECHNICAL SUPPORT ADVISED THAT THE FLAT DETECTOR COOL AND PUMP WAS OUT AND IN NEED OF REPLACEMENT. TECH. SUPPORT ADVISED THAT THE PART WOULD BE ORDERED AND AVAILABLE THE NEXT DAY AT NOON. AT NO TIME DID TECH SUPPORT ADVISE THE TEAM LEAD THAT THE SYSTEM WOULD BECOME INOPERABLE. THE GE LAB THAT WAS REPLACED WOULD OVERHEAD ROUTINELY BUT REMAIN OPERABLE, SO STAFF WERE NOT CONCERNED THAT THE SYSTEM WOULD SHUT DOWN. INTERVENTIONAL CASED STARTED BY THE PHYSICIAN. ENGAGED RIGHT CORONARY ARTERY WITH THE GUIDING CATHETER, ONCE INSERTED, PT EXPRERIENCING CHEST PAIN, AGITATION, DIAPHORESIS AND MILDLY COMBATIVE. PT TREATED WITH MEDICATIONS. BALLOON INSERTED IN PRXIOMAL RIGHT CORONARY AND INFLATED FOR 30 SECONDS. FLOURO FIALED, SYSTEM REBOOTED AND CAME BACK UP FOR 30 SECONDS. SYSTEM FAILED AGAIN AND COMPLETELY SHUT DOWN. REMOVAL OF THE BALLOON, GUIDEWIRE, AND SHEATH SEWIN IN PLACE. PT STILL AGITIATED AND COMBATIVE REQUIRED 4-5 STAFF TO HOLD PT ON STRETCHER. PT TAKEN TO ICU."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLURA XPER FD10 ANGIOGRAPHIC X-RAY SYSTEM IZI PHILIPS MEDICAL SYSTEMS ALLURA XPER FD10 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention