FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 5976777 · Received September 26, 2016

Report

Report Number
1000113657-2016-01635
Event Type
Malfunction
Date Received
September 26, 2016
Date of Event
September 4, 2016
Report Date
November 21, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT RETURNED FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER IS TESTING WITH EXPIRED TEST STRIPS. VIAL LOT OPENED OVER SIX MONTHS.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT SYSTEM EVALUATED WITH NO DEFECT FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIPS HAD A POOR FILL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER'S SON, IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM ALL OF THE RESULTS OBTAINED. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 90 TO 100 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/15/2017 AND OPEN VIAL DATE IS OPEN OVER SIX MONTH. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). THE CUSTOMER HAS NOT HAD ANY CHANGES IN THE DIET. CUSTOMER STATED THAT SHARES THE METER WITH THE WIFE. CUSTOMER'S SON WAS ADVISE THAT THE FATHER SHOULD NOT SHARE THE METER WITH ANYONE. UNABLE TO PERFORM A BACK TO BACK BLOOD TEST AS THE TEST STRIPS HAS BEEN OPEN OVER 6 MONTHS. THE CUSTOMER WAS ADVISED THE TEST STRIPS ARE ONLY GOOD FOR 4 MONTHS AFTER BEING OPEN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER'S SON, IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM ALL OF THE RESULTS OBTAINED. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 90 TO 100 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/15/2017 AND OPEN VIAL DATE IS OPEN OVER SIX MONTH. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). THE CUSTOMER HAS NOT HAD ANY CHANGES IN THE DIET. CUSTOMER STATED THAT SHARES THE METER WITH THE WIFE. CUSTOMER'S SON WAS ADVISED THAT THE FATHER SHOULD NOT SHARE THE METER WITH ANYONE. UNABLE TO PERFORM A BACK TO BACK BLOOD TEST AS THE TEST STRIPS HAS BEEN OPEN OVER 6 MONTHS. THE CUSTOMER WAS ADVISED THE TEST STRIPS ARE ONLY GOOD FOR 4 MONTHS AFTER BEING OPEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
628701 TRUERESULT BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUERESULT PR2069

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY