FDA Adverse Event Malfunction Summary report: N

VITEK® 2 GRAM-NEGAITIVE ANTIMICROBIAL SUSCEPTIBILITY TEST AST-GN73 TEST KIT

MDR report key: 5973507 · Received September 23, 2016

Report

Report Number
1950204-2016-00135
Event Type
Malfunction
Date Received
September 23, 2016
Report Date
February 7, 2017
Manufacturer
BIOMERIEUX, INC
Product Code
LON
PMA / PMN Number
N50510: S130
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE NOT RETURNED TO MANUFACTURER.

Additional Manufacturer Narrative · 1

A BIOMÉRIEUX INTERNAL INVESTIGATION WAS CONDUCTED; HOWEVER, THE CUSTOMER DID NOT COMPLY WITH THE BIOMÉRIEUX REQUEST FOR RAW DATA AND PATIENT ISOLATE SUBMITTAL. THE INVESTIGATION REGARDING A FALSE SUSCEPTIBLE CEFEPIME RESULT FOR ESCHERICHIA COLI, COULD NOT BE COMPLETED, AS ASSESSMENT OF ORGANISM GROWTH IN THE VITEK 2 AST-GN73 CARD CANNOT BE MADE WITHOUT THE RAW DATA OR THE ISOLATE SUBMITTAL. ISOLATE SUBMITTAL IS REQUIRED TO DETERMINE WHETHER THE CEFEPIME RESULT IS DISCREPANT COMPARED TO THE REFERENCE METHOD. A REVIEW OF QUALITY RECORDS CONFIRMED THAT AST-GN73 LOT 5930012403 MET FINAL RELEASE CRITERIA AND NO ISSUES WERE OBSERVED ON INITIAL QC PERFORMANCE TESTING. NO FURTHER INVESTIGATION IS POSSIBLE WITHOUT THE DATA OR ISOLATE.

Description of Event or Problem · 1

A CUSTOMER IN THE UNITED STATES CONTACTED BIOMÉRIEUX TO REPORT THE OCCURRENCE OF A FALSE SUSCEPTIBLE CEFEPIME RESULT FOR ESCHERICHIA COLI IN ASSOCIATION WITH THE VITEK® 2 AST-GN73 TEST KIT. THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMÉRIEUX THAT THE DISCREPANT RESULT LED TO ANY ADVERSE IMPACT TO THE PATIENT'S STATE OF HEALTH. CULTURE SUBMITTALS HAVE BEEN REQUESTED BY BIOMÉRIEUX FOR INTERNAL INVESTIGATION. AN INTERNAL BIOMÉRIEUX INVESTIGATION WILL BE INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
625587 VITEK® 2 GRAM-NEGAITIVE ANTIMICROBIAL SUSCEPTIBILITY TEST AST-GN73 TEST KIT VITEK® 2 AST-GN73 CARD LON BIOMERIEUX, INC 5930012403

Patients

Seq Age Sex Outcome Treatment
1