FDA Adverse Event
Malfunction
Summary report: N
BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 597252
·
Received June 18, 2004
Report
- Report Number
- 2243072-2004-00293
- Event Type
- Malfunction
- Date Received
- June 18, 2004
- Date of Event
- May 20, 2004
- Report Date
- June 18, 2004
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER REPORTS THE REPLACEMENT METER IS GIVING TESTS RESULTA THAT ARE DIFFERENT. ANALYSIS RUN ON CLARK ERROR GRD USING MEAN AVERAGE READING (242) AS REF. POINT VS BD READINGS (288, 141, 297) YEILDED DATA POINTS IN THE D RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 3064100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |