FDA Adverse Event Malfunction Summary report: N

ANALYTICAL P MODULE

MDR report key: 5970010 · Received September 22, 2016

Report

Report Number
1823260-2016-01438
Event Type
Malfunction
Date Received
September 22, 2016
Date of Event
September 12, 2016
Report Date
December 1, 2016
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K953239
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED IN (B)(6).

Additional Manufacturer Narrative · 1

A SPECIFIC ROOT CAUSE COULD NOT BE DETERMINED. AS BOTH RESULTS DIFFERED BY THE SAME AMOUNT AND THE REACTION CURVES WERE ACCEPTABLE, A CALIBRATION ISSUE WAS SUSPECTED.THE FIELD SERVICE REPRESENTATIVE FLUSHED THE SYSTEM AND PERFORMED CHECKS WHICH WERE OK. CALIBRATIONS WERE OK AND PRECISION TESTING WAS PERFORMED. THE ISSUE APPEARED TO BE RESOLVED AFTER THE SERVICE VISIT.

Description of Event or Problem · 1

THE CUSTOMER RECEIVED QUESTIONABLE LOW CREATININE RESULTS FOR 216 PATIENT SAMPLES FROM (B)(6) 2016. DATA WAS ONLY PROVIDED FOR TWO PATIENT SAMPLES THAT OCCURRED ON (B)(6) 2016. PATIENT 1 INITIAL RESULT WAS 122.9 UMOL/L. THE REPEAT RESULTS WERE 191.3 UMOL/L AND 186.9 UMOL/L. PATIENT 2 INITIAL RESULT WAS 70.1 UMOL/L. THE REPEAT RESULTS WERE 111.2 UMOL/L AND 115.3 UMOL/L. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY TO A PHYSICIAN. THE PATIENTS WERE NOT ADVERSELY AFFECTED. THE REAGENT LOT NUMBER WAS 160571. THE EXPIRATION DATE WAS REQUESTED, BUT WAS NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE EXCHANGED THE TURNING TABLE SHEET KIT, CLEANED ALL PIPETTORS, CHECKED THE PRESSURE AND VACUUM, CLEANED THE WATERTANK AND RINSED THE SYSTEM FROM THE WATERTANK. AFTERWARDS, A CELL BLANK WAS OK. THE INVESTIGATION NOTED THE REACTION CURVE LOOKED LIKE WATER DROPLETS HAD FALLEN IN THE CELL DURING THE REACTION. WITH THE PROVIDED INFORMATION AND PRECISION RESULTS, AN ISSUE WITH REAGENT PIPETTING WAS EXCLUDED.