FDA Adverse Event Injury Summary report: N

BONE PIN, 3.2MM X 80MM, STERILE 2 PACK

MDR report key: 5969064 · Received September 21, 2016

Report

Report Number
3005985723-2016-00300
Event Type
Injury
Date Received
September 21, 2016
Date of Event
September 8, 2016
Report Date
September 8, 2016
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K142530
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT: 3.2X80 MM BONE PINS BROKE WHEN BEING DRILLED INTO TIBIA. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION WAS NOT PERFORMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. DIMENSION INSPECTION COMPLETED AT THE TIME OF MANUFACTURING SHOWS THE LOT WAS WITHIN SPECIFICATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (143000-01 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED SHIPPED TO MILLSTONE ON MARCH 29, 2016 AND ACCEPTED INTO FINAL STOCK ON MARCH 29, 2016 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 143080, LOT NUMBER W44718 SHOWS THREE COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. THE RELATED PI NUMBERS ARE 1276313, 1302663, AND 1348408. TRACKING OF COMPLAINTS RELATED TO THE 143080 PART NUMBER WILL BE TRACKED THROUGH QUARTERLY TREND REQUEST #789.. CONCLUSIONS: AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED. IT WAS DETERMINED THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING PLACEMENT OF THE BONE PIN. DUE TO THE NON-USE OF A DRILL GUIDE AS REQUIRED PER MAKOPLASTY PARTIAL KNEE APPLICATION USER GUIDE, 206388 REVISION 01 FOR PLACEMENT OF BONE PINS, THIS EVENT CONSTITUTES AN OFF-LABEL USE OF THE DEVICE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD. THE DEVICE WAS NOT RETURNED FO EVALUATION.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

A SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO). DURING THE CASE WHILE THE SURGEON WAS DRILLING A BONE PIN INTO THE PATIENT'S TIBIA, THE BONE PIN BROKE OFF INSIDE OF THE TIBIA. THE BONE PIN WAS LEFT INSIDE OF THE PATIENT AND THE OUTCOME OF THE CASE WAS SUCCESSFUL.

Description of Event or Problem · 1

A SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO). DURING THE CASE WHILE THE SURGEON WAS DRILLING A BONE PIN INTO THE PATIENT'S TIBIA, THE BONE PIN BROKE OFF INSIDE OF THE TIBIA. THE BONE PIN WAS LEFT INSIDE OF THE PATIENT AND THE OUTCOME OF THE CASE WAS SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
619261 BONE PIN, 3.2MM X 80MM, STERILE 2 PACK STEREOTAXIC DEVICE, ACCESSORY OLO MAKO SURGICAL CORP. W44718-2

Patients

Seq Age Sex Outcome Treatment
1 Other