FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 5967679 · Received September 21, 2016

Report

Report Number
2031642-2016-02421
Event Type
Malfunction
Date Received
September 21, 2016
Report Date
August 30, 2016
Manufacturer
RESPIRONICS CALIFORNIA, INC
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE POWER MANAGEMENT BOARD WAS RETURNED TO THE MANUFACTURER FOR FAILURE INVESTIGATION. VISUAL INSPECTION OF THE POWER MANAGEMENT BOARD (PM PCBA) REVEALED NO EVIDENCE OF DAMAGE OR CONTAMINATION, DURING DEBUGGING THE FAILURE INVESTIGATOR FOUND Y1 (10 MHZ OSCILLATOR) ON THE PM PCBA WITH NO 10 MHZ CLK OUTPUT SIGNAL AND Q42 (N-CHANNEL LOGIC LEVEL ENHANCEMENT MODE FIELD EFFECT TRANSISTOR) PIN 1 DRAIN TO PIN2 GATE SHORTED. IT WAS DETERMINED THAT THE COMPONENT FAILURE Y1 AND Q42 SHORTED ON THE PM PCBA PREVENTED THE VENTILATOR FROM POWERING ON.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE VENTILATOR WILL NOT POWER ON EITHER ON AC OR DC POWER. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT.