FDA Adverse Event
Malfunction
Summary report: N
V60 VENTILATOR
MDR report key: 5967679
·
Received September 21, 2016
Report
- Report Number
- 2031642-2016-02421
- Event Type
- Malfunction
- Date Received
- September 21, 2016
- Report Date
- August 30, 2016
- Manufacturer
- RESPIRONICS CALIFORNIA, INC
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE POWER MANAGEMENT BOARD WAS RETURNED TO THE MANUFACTURER FOR FAILURE INVESTIGATION. VISUAL INSPECTION OF THE POWER MANAGEMENT BOARD (PM PCBA) REVEALED NO EVIDENCE OF DAMAGE OR CONTAMINATION, DURING DEBUGGING THE FAILURE INVESTIGATOR FOUND Y1 (10 MHZ OSCILLATOR) ON THE PM PCBA WITH NO 10 MHZ CLK OUTPUT SIGNAL AND Q42 (N-CHANNEL LOGIC LEVEL ENHANCEMENT MODE FIELD EFFECT TRANSISTOR) PIN 1 DRAIN TO PIN2 GATE SHORTED. IT WAS DETERMINED THAT THE COMPONENT FAILURE Y1 AND Q42 SHORTED ON THE PM PCBA PREVENTED THE VENTILATOR FROM POWERING ON.
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE VENTILATOR WILL NOT POWER ON EITHER ON AC OR DC POWER. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT.