FDA Adverse Event Malfunction Summary report: N

ONCOR IMPRESSION PLUS

MDR report key: 5966204 · Received September 21, 2016

Report

Report Number
2240869-2016-52796
Event Type
Malfunction
Date Received
September 21, 2016
Date of Event
August 25, 2016
Report Date
August 26, 2016
Manufacturer
SIEMENS HEALTHCARE GMBH
Product Code
IYE
PMA / PMN Number
K060226
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE REPORTED EVENT IS ON-GOING. EMERGENCY STOP BUTTONS ARE AVAILABLE ON THE UNIT TO STOP ALL SYSTEM MOVEMENT. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WAS SUBMITTED SEPTEMBER 21, 2016.

Description of Event or Problem · 1

THE CUSTOMER HAD A PAUSE ENTERED BETWEEN THE SECOND SUPERIOR ANTERIOR OBLIQUE (SAO) FIELD AND THE POSTERIOR-ANTERIOR (PA) FIELD TO PREVENT AUTOMATIC GANTRY ROTATION IN ORDER TO AVOID ANY COLLISION WITH PATIENT AND TABLE. THE PATIENT WAS DUE FOR A FILM OF THE PA FIELD. DUE TO RISK OF COLLISION THE THERAPISTS PRESENT AT THE TREATMENT DISABLED THE AUTO DEPLOY ON THE FLAT PANEL. WHEN THE THERAPISTS DOWNLOADED THE TWO FIELDS THERE WAS A STOP/PAUSE ON THE PA FIELD AS USUAL. WHEN THE TREATMENT OF THE SECOND SAO FIELD WAS COMPLETED, THE THERAPISTS NOTICED THAT THE GANTRY STARTED ROTATING ON ITS OWN, EVEN WITH A PAUSE STILL BEING DISPLAYED. THE THERAPISTS HIT THE RAD OFF BUTTON AND THE BUTTON TO OPEN THE DOOR BEFORE THE GANTRY HIT THE PATIENT'S ARMS, THE VAC-LOC AND THE TABLE. THE GANTRY MOVEMENT WAS STOPPED. IT IS ASSUMED THAT HAVING A FILM IN THE SESSION DISABLED THE PAUSE THAT HAD BEEN ENTERED EARLIER, HOWEVER, THE PAUSE WAS STILL DISPLAYED DUE THE AUTO-DEPLOY BEING DISABLED. THERE ARE NO INJURIES ATTRIBUTED TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
619783 ONCOR IMPRESSION PLUS MEDICAL CHARGED-PARTICLE RADIATION THERAPY SYSTEM IYE SIEMENS HEALTHCARE GMBH 5857912

Patients

Seq Age Sex Outcome Treatment
1 Other