ONCOR IMPRESSION PLUS
Report
- Report Number
- 2240869-2016-52796
- Event Type
- Malfunction
- Date Received
- September 21, 2016
- Date of Event
- August 25, 2016
- Report Date
- August 26, 2016
- Manufacturer
- SIEMENS HEALTHCARE GMBH
- Product Code
- IYE
- PMA / PMN Number
- K060226
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION OF THE REPORTED EVENT IS ON-GOING. EMERGENCY STOP BUTTONS ARE AVAILABLE ON THE UNIT TO STOP ALL SYSTEM MOVEMENT. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WAS SUBMITTED SEPTEMBER 21, 2016.
THE CUSTOMER HAD A PAUSE ENTERED BETWEEN THE SECOND SUPERIOR ANTERIOR OBLIQUE (SAO) FIELD AND THE POSTERIOR-ANTERIOR (PA) FIELD TO PREVENT AUTOMATIC GANTRY ROTATION IN ORDER TO AVOID ANY COLLISION WITH PATIENT AND TABLE. THE PATIENT WAS DUE FOR A FILM OF THE PA FIELD. DUE TO RISK OF COLLISION THE THERAPISTS PRESENT AT THE TREATMENT DISABLED THE AUTO DEPLOY ON THE FLAT PANEL. WHEN THE THERAPISTS DOWNLOADED THE TWO FIELDS THERE WAS A STOP/PAUSE ON THE PA FIELD AS USUAL. WHEN THE TREATMENT OF THE SECOND SAO FIELD WAS COMPLETED, THE THERAPISTS NOTICED THAT THE GANTRY STARTED ROTATING ON ITS OWN, EVEN WITH A PAUSE STILL BEING DISPLAYED. THE THERAPISTS HIT THE RAD OFF BUTTON AND THE BUTTON TO OPEN THE DOOR BEFORE THE GANTRY HIT THE PATIENT'S ARMS, THE VAC-LOC AND THE TABLE. THE GANTRY MOVEMENT WAS STOPPED. IT IS ASSUMED THAT HAVING A FILM IN THE SESSION DISABLED THE PAUSE THAT HAD BEEN ENTERED EARLIER, HOWEVER, THE PAUSE WAS STILL DISPLAYED DUE THE AUTO-DEPLOY BEING DISABLED. THERE ARE NO INJURIES ATTRIBUTED TO THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619783 | ONCOR IMPRESSION PLUS | MEDICAL CHARGED-PARTICLE RADIATION THERAPY SYSTEM | IYE | SIEMENS HEALTHCARE GMBH | 5857912 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |