TRUEMETRIX
Report
- Report Number
- 1000113657-2016-01584
- Event Type
- Malfunction
- Date Received
- September 20, 2016
- Date of Event
- August 29, 2016
- Report Date
- October 3, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI
- Reporter Occupation
- OTHER
Narratives
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND ON RETURNED METER AND RETURNED TEST STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
(B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER HAD AN INACCURATE REFERENCE.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 113, 133 AND 158 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 89 - 109 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/15/2017 AND OPEN VIAL DATE IS 08/22/2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 113MG/DL (B)(6) 2016 10:16 AM FASTING: YES; 133MG/DL (B)(6) 2016 11:09 AM FASTING: YES; 158MG/DL (B)(6) 2016 09:28 AM FASTING: YES; 115MG/DL (B)(6) 2016 02:16 PM FASTING: YES; 191MG/DL (B)(6) 2016 11:41 AM FASTING: YES. MEMORY CONCERNS: 113 133 158 115 191MG/DL.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 113, 133 AND 158 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 89 - 109 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/15/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 1: 113MG/DL (B)(6) 2016 10:16 AM FASTING: YES, 2: 133MG/DL (B)(6) 2016 11:09 AM FASTING: YES, 3: 158MG/DL (B)(6) 2016 09:28 AM FASTING: YES, 4: 115MG/DL (B)(6) 2016 02:16 PM FASTING: YES, 5: 191MG/DL (B)(6) 2016 11:41 AM FASTING: YES, MEMORY CONCERNS: 113 133 158 115 191MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 615835 | TRUEMETRIX | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUEMETRIX | MT1845 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |