FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 5963302 · Received September 20, 2016

Report

Report Number
1000113657-2016-01584
Event Type
Malfunction
Date Received
September 20, 2016
Date of Event
August 29, 2016
Report Date
October 3, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND ON RETURNED METER AND RETURNED TEST STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 113, 133 AND 158 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 89 - 109 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/15/2017 AND OPEN VIAL DATE IS 08/22/2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 113MG/DL (B)(6) 2016 10:16 AM FASTING: YES; 133MG/DL (B)(6) 2016 11:09 AM FASTING: YES; 158MG/DL (B)(6) 2016 09:28 AM FASTING: YES; 115MG/DL (B)(6) 2016 02:16 PM FASTING: YES; 191MG/DL (B)(6) 2016 11:41 AM FASTING: YES. MEMORY CONCERNS: 113 133 158 115 191MG/DL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 113, 133 AND 158 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 89 - 109 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/15/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 1: 113MG/DL (B)(6) 2016 10:16 AM FASTING: YES, 2: 133MG/DL (B)(6) 2016 11:09 AM FASTING: YES, 3: 158MG/DL (B)(6) 2016 09:28 AM FASTING: YES, 4: 115MG/DL (B)(6) 2016 02:16 PM FASTING: YES, 5: 191MG/DL (B)(6) 2016 11:41 AM FASTING: YES, MEMORY CONCERNS: 113 133 158 115 191MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
615835 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1845

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY