Summary report: N
BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 596260
·
Received April 26, 2005
Report
- Report Number
- 2243072-2005-00079
- Event Type
- Other
- Date Received
- April 26, 2005
- Date of Event
- February 14, 2005
- Report Date
- March 11, 2005
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT GETTING HIGH READINGS WITH THEIR PINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (353) AS REF. POINT VS. BD READINGS OF 464, 339, 233, 375 YIELDING DATA POINTS IN THE C ZONE OF GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 4064183 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |