SET SCREW (TI-6AL-4V ELI)
Report
- Report Number
- 2027467-2016-00070
- Event Type
- Injury
- Date Received
- September 19, 2016
- Date of Event
- August 8, 2016
- Report Date
- August 16, 2016
- Manufacturer
- ALPHATEC SPINE INC
- Product Code
- NKB
- UDI-DI
- 00840967124582
- PMA / PMN Number
- K133221
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
A REVIEW OF THE RETURNED SET SCREWS REVEALED MINOR THREAD PROFILE IRREGULARITIES. TO HELP AID IN DETERMINING ROOT CAUSE, THE SURGEON VISITED ALPHATEC ON (B)(6) 2016. HE INDICATED THAT ALTHOUGH CROSS-THREADING OCCURS DURING HIS CASES PERIODICALLY, THIS PARTICULAR CASE WAS CONCERNING BECAUSE IT APPEARED TO HAPPEN REPEATEDLY. THE SURGEON TOOK PART IN A BLIND EVALUATION IN WHICH SET SCREWS WERE INSERTED AT VARIOUS CROSS-THREADED ANGLES TO SEE IF A DIFFERENCE IN FEEL/PERFORMANCE COULD BE DETERMINED. DURING THIS EVALUATION, IT WAS NOTICED THAT THE SURGEON USES A NON-STANDARD SET SCREW INSERTION TECHNIQUE, WHICH WHEN COMBINED WITH THE DESIGN OF THE ARSENAL SET SCREWS WITH A LARGE LEAD-IN CHAMFER AND A MINIMIZED PROFILE, MAY INCREASE THE CHANCE OF CROSS-THREADING.
THERE WERE A TOTAL OF SIX (6) ARSENAL SET SCREWS RETURNED FOR EVALUATION. ALL SIX (6) ARE THE SAME PRODUCT PART NUMBER/DESCRIPTION BUT CONTAIN FOUR (4) UNIQUE IDENTIFYING LOT NUMBERS. LOT 700941 (2 PIECES) MANUFACTURED 12/8/2015. LOT 7766301 (2 PIECES) MANUFACTURED 5/6/2016. LOT 696254 (1 PIECE) MANUFACTURED 9/14/2015. LOT 700611 (1 PIECE) MANUFACTURED 12/8/2015. THE ARSENAL SET SCREWS ARE CURRENTLY BEING EVALUATED. A FOLLOW UP REPORT WITH RESULTS OF THE INVESTIGATION WILL BE SUBMITTED UPON COMPLETION.
SURGEON HAD AN ISSUE WITH INTRA-OPERATIVE CROSS-THREADING/INITIATION OF SET SCREWS. THE EVENT CAUSED OVER A 30 MINUTE DELAY IN THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612274 | SET SCREW (TI-6AL-4V ELI) | PEDICLE SCREW SYSTEM | NKB | ALPHATEC SPINE INC | 47127 | 00840967124582 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Other |