FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 5960230 · Received September 19, 2016

Report

Report Number
1000113657-2016-01572
Event Type
Malfunction
Date Received
September 19, 2016
Date of Event
August 26, 2016
Report Date
October 13, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECT FOUND ON RETURNED METER AND RETURNED TEST STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

COSTUMER REPORTED COMPLAINT FOR HIGH TEST RESULTS. MOTHER IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM BACK TO BACK BLOOD TEST OF 221, 209 AND 161 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 85-150 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 7/31/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). CUSTOMER IS AT SCHOOL. UNABLE TO VERIFY IF HE IS SYMPTOMATIC. CUSTOMER'S MOTHER WAS TRAINED ON PROPER BACK TO BACK TESTING TECHNIQUE. RESULT FROM CONTROL SOLUTION IS WITHIN NORMAL RANGE.

Description of Event or Problem · 1

COSTUMER REPORTED COMPLAINT FOR HIGH TEST RESULTS. MOTHER IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM BACK TO BACK BLOOD TEST OF 221, 209 AND 161 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 85-150 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 7/31/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). CUSTOMER IS AT SCHOOL. UNABLE TO VERIFY IF HE IS SYMPTOMATIC. CUSTOMER'S MOTHER WAS TRAINED ON PROPER BACK TO BACK TESTING TECHNIQUE. RESULT FROM CONTROL SOLUTION IS WITHIN NORMAL RANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611228 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1573

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY