VERSA-DIAL SHOULDER SYSTEM MODULAR HEAD W/VARIABLE OFFSET 42MM X 18MM X 46MM
Report
- Report Number
- 0001825034-2016-03656
- Event Type
- Injury
- Date Received
- September 16, 2016
- Date of Event
- March 12, 2015
- Report Date
- May 25, 2017
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- MBF
- PMA / PMN Number
- PK060716
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 13 STATES, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN."
CONCOMITANT MEDICAL PRODUCT: COMPREHENSIVE PRIMARY STEM 11 MM MINI, CATALOG#: 113631, LOT#: 503990; VERSA-DIAL/COMPREHENSIVE TI STANDARD TAPER, CATALOG#: 118001, LOT#: 115820; MD HYBRID GLENOID BASE 4 MM, CATALOG#: 113954, LOT#: 696930; PT HYBRID GLENOID POST REGENEREX, CATALOG#: PT-113950, LOT#: 052420. REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. X-RAY REVIEW INDICATES THAT PROGRESSIVE GLENOID COMPONENT RADIOLUCENCY IS DETECTED IN ZONES 1, 2, 3, AND 4 BETWEEN THE MONTH 3 AND MONTH 12 FOLLOW-UP VISITS. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION(S) IS/ARE REQUIRED. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND CORRECTED INFORMATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-02414/ 0001825034-2017-02416/ 0001825034-2017-02417/ 0001825034-2017-02418/0001825034-2017-02419/0001825034-2017-02420/0001825034-2017-02421/0001825034-2017-02422/0001825034-2017-02423.
PATIENT ENROLLED IN A CLINICAL STUDY REPORTED RIGHT SHOULDER PAIN AND PULLING SENSATION AROUND THE INCISION SITE APPROXIMATELY 1 YEAR POST-IMPLANTATION.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT TOTAL SHOULDER ARTHROPLASTY. SUBSEQUENTLY, PAIN, INSTABILITY, AND ACROMIOCLAVICULAR JOINT TENDERNESS WERE NOTED AT THREE MONTH POST-OPERATIVE FOLLOW-UP. AT ONE YEAR POST-OPERATIVE FOLLOW-UP, BICEPS TENDON TENDERNESS, IMPINGEMENT, UNUSUAL SHOULDER PAIN AND PULLING SENSATION AROUND THE INCISION SITE, AND A "PUCKERING IN" DEFORMITY OF THE RIGHT MIDDLE OF THE SHOULDER WERE REPORTED. INSTABILITY HAD RESOLVED. NO REVISION PROCEDURE HAS BEEN INDICATED AT THIS TIME, AS THE ISSUE REMAINS TOLERATED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT TOTAL SHOULDER ARTHROPLASTY. SUBSEQUENTLY, PAIN AND ACROMIOCLAVICULAR JOINT TENDERNESS WERE NOTED AT THREE MONTH POST-OPERATIVE FOLLOW-UP. AT ONE YEAR POST-OPERATIVE FOLLOW-UP, BICEPS TENDON TENDERNESS, IMPINGEMENT, UNUSUAL SHOULDER PAIN AND PULLING SENSATION AROUND THE INCISION SITE, AND A "PUCKERING IN" DEFORMITY OF THE RIGHT MIDDLE OF THE SHOULDER WERE REPORTED. NO REVISION PROCEDURE HAS BEEN INDICATED AT THIS TIME, AS THE ISSUE REMAINS TOLERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 609046 | VERSA-DIAL SHOULDER SYSTEM MODULAR HEAD W/VARIABLE OFFSET 42MM X 18MM X 46MM | PROSTHESIS, SHOULDER | MBF | BIOMET ORTHOPEDICS | N/A | 620590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Other |